Vacuum Oven for Antibiotic Experiment

 

VO-640S, VO-847S 


Purpose 

  • Antibiotic testing and drying
  • Vacuum oven for pharmaceutical & biotech industries
  • This vacuum oven is useful for antibiotic testing, drying, and similar applications in the biotech industry. The temperature range is adjustable depending on the intended use, and the internal materials and heating methods can be appropriately applied based on the temperature range.
  • Options such as heater capacity, temperature, and test range can be selected based on the conditions of use, with consultation available for decision-making. (In accordance with industrial safety and health regulations, we would like to inform you that IREA TECH products are designed and manufactured according to KS-6750.)


Feature

Stable Vacuum Maintenance & Safe Door Structure

  • PID Auto-tuning Programmable Controller
    Auto-tuning : The PID values are automatically adjusted according to the surrounding environment, enabling precise temperature control (Touchscreen Monitor option). 
  • Confuguration
    ① The oven is highly durable, even for drying solvents and corrosive substances.
    ② The corners are designed in a round shape, making cleaning easy.
    ③ The shelves are made of stainless steel, offering excellent heat transfer, durability, and corrosion resistance.
    ④ The door has a double structure with reinforced glass, allowing visibility of the interior.
    ⑤ The glass door design minimizes resistance when closing due to the packing, ensuring easy closure and a secure vacuum seal. (Easy interior observation)
    ⑥ The packing is made of high-temperature grade silicon material, providing excellent sealing performance (can maintain up to 0.1 torr at 20°C ambient temperature).
    ⑦ The vacuum line and vacuum release line are separated, and the vacuum line has a digital vacuum fitting, allowing quick and precise vacuum settings.
    ⑧ The use of a high-quality valve allows for adjustable vacuum release speed, gradually introducing air to prevent powder sample dispersion. The structure prevents damage to the valve when air is introduced and discharged.
    ⑨ A vacuum oil backflow prevention solenoid valve is installed to prevent contamination of the chamber interior by vacuum oil (Standard).
    ⑩ A digital vacuum controller and Pirani gauge enable setting the vacuum level in cycles (Option).
    ⑪ The over-temperature limit system automatically cuts off power in case of overheating, ensuring fire prevention and safe operation. 




Specification

Chamber Volume (ℓ)640L847L 
TemperatureRange 50℃~150℃
Heat-up Time100℃ within 60 min
Controller Digital PID Controller 
Vacuum GaugeGauge Analogue or Digital Gauge 
MaterialChamber Stainless Steel 304
BodySteel, 1.2t, Power Coating
Sensor K-Type
InsulationCeramic Wool
DoorSUS 304 25t
Door GasketHeat Resistance Silicone (O-ring Type) 
Viewing WindowTemperature safety Glass, 15t
ValveVacuum line Pneumatic Type Angle Valve / NW-40
Vent lineManual Type 3/8" needle valve with filter
HeaterHeating TypeSUS SHEATH Heater
Capacity 12~16 kW
Size (W x D x H,  mm)Internal 800 x 980 x 800900 x 960 x 980 
External1100 x 1280 x 12001200 x 1230 x 1800 
Power Consumption220V 380V 3P, 50/60Hz
Safety DeviceOver-Temperature Buzzer, Over-Temp. Limit, Circuit Breaker 

※ Size and other specifications are always negotiable with clients' demand.
※ IQ.OQ is conducted according to the customer's request. 




Vacuum range options 

10-3Torr or higher is also available.